Low moisture nội dung measurement at 10 ppm (0.001%) readability (MS-70) Fast and uniform heating with halogen lamp & innovative Secondary Radiation Assist (SRA) technology Accuracy verification with Sodium Tartrate Dihydrate (standard with MS-70, MX-50, & MF-50) High repeatability with the Super Hybrid Sensor (SHS) technology Calibration function for heating temperature (for MS-70 and MX-50 only) Five different measurement programs Selectable heating patterns Translucent window enabling progress kiểm tra Memory function for measurement conditions Large bright Vacuum Fluorescent Display GLP/GMP/GCP/ISO compliant đầu ra Quick reference card provided
Videos
Windows Communication Tools WinCT-Moisture (standard with MS-70 & MX-50)
Displays moisture rate changes over time in a graph (RsFig) Quickly & automatically suggests the most suitable heating conditions for first-time samples (RsTemp) Calculates measurement data statistics Saves the recorded data in CSV format Detects other changes khổng lồ the sample material in addition khổng lồ moisture rate Includes sample data summary USB connection is also possible (AX-USB-25P)
External Input/Output
RS-232C interface
Accessories
AX-MX-30 : Disposable aluminum pan (100 pcs)AX-MX-31 : Sample pan (100 pcs)AX-MX-32-1 : Filtration glass fiber sheet (100 sheets)AX-MX-32-2 : Glass fiber sheet (100 sheets)Effective Use of Various Aluminum Pans / Glass-Fiber Sheets (PDF)AX-MX-33 : Sodium Tartrate Dihydrate (30 g × 12 pcs) Material Safety Data Sheet (PDF)AX-MX-34-120V : Halogen lamp (AC 100 to 120 V)AX-MX-34-240V : Halogen lamp (AC 200 to 240 V)AX-MX-35 : Pan handle (2 pcs)AX-MX-36 : Tweezers (2 pairs)AX-MX-37 : Spoon (2 pcs)AX-MX-38 : Display cover (2 pcs)AX-MX-39 : Dust coverAX-MX-40 : RS-232C cable (25P-9P, 2 m)AX-MX-41 : Calibration weight (20 g)AX-MX-42 : WinCT-MoistureAX-MX-43 : Temperature calibrator (for MS-70 and MX-50 only)Manual for AX-MX-43 (PDF)AX-USB-25P : RS-232C/USB converter with cableAD-1683 : Static eliminatorAD-1684A : Electrostatic field meter
Specifications
Model
MS-70
MX-50
MF-50
ML-50
Measurement method
400 W straight halogen lamp heating system with SRA filter & SHS weighing technology
Maximum sample weight capacity
71 g
51 g
Readability
Moisture content
0.001%/0.01%/0.1%
0.01%/0.1%
0.05%/0.1%/1%
0.1%/1%
Weight
0.0001 g
0.001 g
0.002 g
0.005 g
Repeatability(Standard deviation)
Moisture nội dung (over 1 g)
0.05%
0.1%
0.2%
0.5%
Moisture content (over 5 g)
0.01%
0.02%
0.05%
0.1%
Weight
0.0005 g
0.001 g
0.002 g
0.005 g
Drying temperature
30 lớn 200 C°
50 lớn 200 C°
Measurement unit
Wet base / Dry base / Dry content / Ratio / Weight
Sample pan size
Ø85 mm
Note: Opening/closing the heater cover.
This instrument is a precision instrument. Open/close the heater cover gently when placing/removing a sample at the start/end of a measurement.